Welcome to the Serik Speed fire calibration tutorial. Calibration is required after replacing the base plate or the thermo couple, and in order to examine suspected errors and correct potential drift. During this process, the values from the internal thermo couple is compared to an external thermo couple. The values collected from the external thermal couple are then applied to the furnace. It is important that the unit is at room temperature for the calibration process. The calibration kit comes with the following components. A calibration tool, the top door insulation with continuous hole drilled through and an alignment tool for the thermo couple. Use caution when handling the white door insulation material. Welcome to the in lab M C. X five Maintenance protocol. This video will cover the necessary measures to successfully complete than EMC X five yearly maintenance prior to starting the maintenance ensure from the lab that the unit is and has been working properly informed the staff. If there are any found defects as these will not be covered by the maintenance, print out the in lab M c X five maintenance certificate from dense supply. Sarona support dot com. The required tools will be as follows a torques 20 driver, a torques 10 driver, an explorer and a Phillips head screwdriver. The maintenance kit comes packaged with the air regulator. Filter the in line water, filter the water tank filter, adore sponge, suction, bellows, a spindle maintenance. Get clamping screws, O rings, host clips and cleaning brushes. If the software is in need of updating schedule up to 2.5 hours for a complete maintenance, we are ready to start the maintenance with a machine powered on. The first component to check is the general condition of the unit at this point.